Turbidity measurement of water samples

                                                   Experiment no. 1

AIM : TO DETERMINE TURBIDITY OF GIVEN WATER SAMPLE.

APPARATUS REQUIRED:  1. Turbidity meter with its accessories.     
                                                2. container , beaker
                                                  3. water sample, soft cotton cloth


THEORY :
             
Turbidity in water is due to suspended particles like clay, clay ,silt,finely divided matters and macroscopic organisms.
                                   Turbidity meter is used for measurement of turbidity in Nephelometer turbidity unit i.e.NTU with reference to standard solutions.
                                                                      Turbidity meter basically involves scattering phenomenon due to suspended particles when light passes through sample. this scattering of light is generally proportional to turbidity.



PROCEDURE:-
                      1. standardise the turbidity meter. select the turbidity change according to expected value of meter.

2. push the calibration key to enter the calibration(calibration is basically satisfying the standards) mode follow instrument instructions on display.

3.insert the 20 NTU standard solution and close the lid .

4.push read button, the display shows stabilizing and than shows the result.

5. repeat step 2&3 with 100  and 800 NTU
to calibrate standards.

6. save the results and press exit , clean the water sample with cotton cloth.

7.place sample cells in turbidity meter and close the lid , press read and turbidity value is measured as NTU.

CALIBRATION TABLE:

S.no.                       Standard                                      turbidity
                              Turbidity                                     obtained
                             Of sample(NTU)                           (NTU)

1.                             10         9.88

2.                            20           20
   
3.                            100         99-101

4.                            800         793-802



OBSERVATION TABLE:-

S.NO.                   Water sample                          Turbidity
                                                                             (NTU)
 
1.                           SAMPLE1   0.51                

2.                           SAMPLE 2    29.5

3.                         TAP WATER   0.95

4.                         POND WATER


PRECAUTIONS:-

1.CALIBERATE the meter and verify the calibration mode.
2.Do not air dry the sample cells.
3.wipe the outside of the sample cells with soft cotton cloth.
4.Reading must be taken carefully.

RESULT:-  ....................................

If turbidity of water  sample is less than 5 NTU then it is safe for drinking purpose, if greater than 5 then not safe.